Endoleak classification
Persistent flow in the aneurysm sac outside the stent graft after EVAR/TEVAR. Types I and III are high pressure and need prompt treatment.
| Type | Source | Imaging clue |
|---|---|---|
| Ia / Ib | Inadequate seal at proximal (Ia) or distal (Ib) attachment site. Ic: iliac occluder | Contrast contiguous with graft end; arterial phase |
| II | Retrograde flow from branch vessels (IMA, lumbar arteries) | Peripheral sac contrast near IMA/lumbar origin; often delayed phase only |
| III | Graft failure: junctional separation (IIIa) or fabric tear (IIIb) | Contrast adjacent to graft body or overlap zone |
| IV | Graft porosity, within 30 days of placement | Diffuse blush on completion angiogram |
| V | Endotension: sac expansion without visible leak | Enlarging sac, no contrast on any phase |
Triple-phase CTA (non-contrast, arterial, delayed) separates endoleak from calcification and catches slow type II leaks. Sac growth >5 mm warrants investigation.